namespace:chiller_flush
Differences
This shows you the differences between two versions of the page.
| Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
| namespace:chiller_flush [2026/02/20 17:33] – admin | namespace:chiller_flush [2026/08/05 16:47] (current) – 12.144.223.107 | ||
|---|---|---|---|
| Line 1: | Line 1: | ||
| =====Inlite Chiller and Recirculator Flush===== | =====Inlite Chiller and Recirculator Flush===== | ||
| + | |||
| + | **LAST REPLACED DI FILTER: July 31st 2026** | ||
| There are times when the Inlite laser head will decrease in energy, as well as needing to replace filters.\\ This should help as a guild so doing these tasks go smoothly and quickly. | There are times when the Inlite laser head will decrease in energy, as well as needing to replace filters.\\ This should help as a guild so doing these tasks go smoothly and quickly. | ||
| Line 10: | Line 12: | ||
| If you want to clean the system. Let a solution of 80% Hydrogen Peroxide and 20% distilled water run overnight. You can leave in Standby mode for this time. | If you want to clean the system. Let a solution of 80% Hydrogen Peroxide and 20% distilled water run overnight. You can leave in Standby mode for this time. | ||
| + | |||
| + | |||
| + | The next morning, pump out all the Solution out and clean with distill water. Then fill the reservoir. | ||
| + | |||
| + | Actually Steps: | ||
| + | |||
| + | - Turn Off the chiller, by the switch and the key | ||
| + | - Open the interlock top cover | ||
| + | - Use the yellow pump to Syphin the water OUT of reservoir | ||
| + | - Unplug the Return line, from the back of the rack, (have 1-2 paper towels ready for a few drops) | ||
| + | - connect the pump to the return hose and syphon the water out of the laser head. | ||
| + | - Fill with 80% Hydrogen peroxide 20% distilled water(2 normal size bottles were enough,) | ||
| + | - you will need to close the interlock and turn on the rack, and pump some water into the lines, and them refill the reservoir a few times | ||
| + | - Let it stay in Standby Mode over night | ||
| + | - Syphon all the solution out | ||
| + | - fill back with Distill water, let it run through the lines, and syphon again. Do this step twice | ||
| + | - replace, the Di-filter, make sure to re tape the threads | ||
| + | - Fill back with distilled water, and fill the lines | ||
| + | - DONE FOR 6months | ||
| + | |||
| + | {{: | ||
| + | |||
| + | {{: | ||
| + | |||
| + | {{: | ||
| + | |||
| + | {{: | ||
namespace/chiller_flush.1771608813.txt.gz · Last modified: by admin